A constant intensity technique to improve the performances of devices based on direct absorption spectroscopy

Year: 2015

Authors: Montori A., De Pas M., Giuntini M., de Cumis MS., Viciani S., D’Amato F.

Autors Affiliation: CNR INO, I-50019 Sesto Fiorentino, FI, Italy; Univ Florence, LENS, I-50019 Sesto Fiorentino, FI, Italy; CNR INO, I-50125 Florence, Italy.

Abstract: We describe an all-in-fibre apparatus for Constant Intensity Direct Absorption Spectroscopy (CIDAS) for gas concentration measurements which keeps the power of a diode laser constant along the frequency sweep. The reduction of the large variation of the laser power, connected to the frequency scan, enhances the ability of detecting small variations in a background signal, resulting in an increase of the sensitivity with respect to standard direct absorption techniques. Moreover, CIDAS allows for a real-time observation of the absorption signals without any kind of post-detection processing. The apparatus has been tested with carbon dioxide (CO2) and methane (CH4), around 1.57 and 1.65 µm, respectively.

Journal/Review: OPTO-ELECTRONICS REVIEW

Volume: 23 (1)      Pages from: 28  to: 32

More Information: The authors acknowledge the financial support of Regione Toscana, in the frame of POR CREO FESR 2007-2013, Project SIMPAS (Innovative Measurement Systems for the protection of Environment and Health).
KeyWords: Carbon dioxide; Fiber optics; Sensitivity analysis, Applied spectroscopy; Background signals; Constant power; Direct absorption; Direct absorption spectroscopies; Gas concentration measurement; Laser-based; Real time observation, Absorption spectroscopy
DOI: 10.1515/oere-2015-0002

ImpactFactor: 1.611
Citations: 1