Note: An analyzer for field detection of H2S by using cavity ring-down at 1.57 mu m
Siciliani de Cumis M., Viciani S., Galli I., Mazzotti D., Sorci F., Severi M., D'Amato F.
INO authors
D’Amato Francesco, Galli Iacopo, Mazzotti Davide, Siciliani de Cumis Mario, Sorci Francesco, Viciani Silvia
Abstract
A prototype analyzer for the detection of hydrogen sulfide (H2S), based on cavity ring-down spectroscopy, is described. The device exploits, whenever possible, optical fibers, in order to simplify the alignment and to improve the stability. A trade-off between low detection level and simplicity has been pursued. The experimental results obtained during tests on different kinds of H2S samples are shown. (C) 2015 AIP Publishing LLC.
Affiliations
CNR, Ist Nazl Ott, I-50019 Sesto Fiorentino, FI, Italy; European Lab Nonlinear Spect LENS, I-50019 Sesto Fiorentino, FI, Italy; CNR, Ist Nazl Ott, I-50125 Florence, FI, Italy; Ecofutura Srl, I-51015 Monsummano Terme, PT, Italy; Univ Florence, Dept Chem Ugo Schiff, I-50019 Sesto Fiorentino, FI, Italy.
More information
The authors acknowledge the financial support of Regione Toscana, in the frame of \"POR CREO FESR 2007-2013,\" Project SIMPAS (Innovative Measurement Systems for the protection of Environment and Health).