Optical Frequency Comb as a general-purpose and wide-band calibration source for astronomical high resolution infrared spectrographs
Schettino G., Oliva E., Inguscio M., Baffa C., Giani E., Tozzi A., Cancio P.
INO authors
Cancio Pastor Pablo, Inguscio Massimo, Schettino Giulia
Abstract
A spectrally filtered Optical Frequency Comb (OFC) laser is proposed as a versatile calibration source for astronomical spectrometers in the 1-2 mu m spectral range. Such a source overcomes the limitations of current calibration lamps providing a uniform spectrum of equally spaced lines with similar intensity and extremely high long-term frequency stability. We present preliminary studies and results of a system which filters the OFC from a 100 MHz comb spacing to 16 GHz one, an adequate spacing for spectrometers with resolving power Delta lambda/lambda > 30000. The first approach employs two Fabry-Perot cavities in series, made of dielectric coated mirrors, followed by a non-linear optical broadening system. The limitations of such a filtering process are discussed. These can be overcome by the second approach, based on filtering cavities with metallic coated mirrors.
Affiliations
CNR - Istituto Nazionale di Ottica, Largo E. Fermi 6, Firenze 50125, Italy;
Dipartimento di Fisica e Astronomia, Via Sansone 1, 50019 Sesto Fiorentino (Fi), Italy
LENS, Via Carrara 1, 50019 Sesto Fiorentino (Fi), Italy;
INAF-Oss. Astrofisico di Arcetri, Largo E. Fermi 5, 50125 Firenze, Italy