Detection of HCl on the first and second overtone using semiconductor diode lasers at 1.7 µm and 1.2 µm

Corsi C., Inguscio M., Chudzynski S., Ernst K., D'Amato F., De Rosa M.

Type
Papers in JCR/ISI journals
Year
1999
Journal
APPLIED PHYSICS B-LASERS AND OPTICS
Volume
68 (2)
Pages
267–269
Impact factor
1.845
Citations
12 (Web of Science, updated on 22 November 2026)
Keywords
Hydrochloric acid; Light absorption; Spectrometers; Semiconductor diode laser spectrometers; Semiconductor lasers

INO authors

Corsi Chiara, De Rosa Maurizio, D’Amato Francesco, Ernst Kcthistov, Inguscio Massimo

Abstract

Using a semiconductor diode laser spectrometer, HCl absorption is studied on the first (2v) and second (3v) overtones at 1.7 micron and 1.2 micron, respectively. The P(3) component at 1207.91 nm and the R(3) component at 1742.38 nm are investigated at different gas pressures, in order to extract the self-broadening parameters.
Sensitivity studies are also performed, to evaluate the minimum detectable concentration of HCl in air.

Affiliations

Dipartimento di Scienze Neurologiche, Università di Firenze, Viale Morgagni 85, 50134 Firenze, Italy; Dipartimento di Fisica, Università di Firenze, Largo E. Fermi 2, 50125 Firenze, Italy; Institute of Experimental Physics, Warsaw University, Warsaw, Poland; ENEA, INN-FIS-SPET, Via E. Fermi 45, 00044 Frascati, Italy; SIT, Via Masaccio 116, Firenze, Italy