Differential absorption imaging for elemental analysis of thin samples using a soft laser-plasma X-ray source

Laville S., Gizzi LA., Köster P., Labate L.

Type
Papers in JCR/ISI journals
Year
2005
Journal
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume
538 (1-3)
Pages
738–746
Impact factor
1.224
Citations
2 (Web of Science, updated on 22 November 2026)
Keywords
synchrotron-radiation; medical applications; emission; angiography; resolution

INO authors

Gizzi Leonida Antonio, Labate Luca

Abstract

The differential X-ray imaging technique is a powerful tool for the detection of small concentrations of contrast agents in samples, as demonstrated recently by many biological and medical applications using synchrotron radiation. In this paper, we present an implementation of this technique using a laser-plasma soft X-ray source combined with an X-ray imaging system based upon a spherically bent Bragg crystal. Using a test sample containing a solution of LiBr of known concentration, we demonstrate that the use of our table-top configuration is suitable to detect Br inside our sample via differential imaging at the L-2 edge of Br at 7.7684 Angstrom. (C) 2004 ElsevierB.V. All rights reserved.

Affiliations

CNR, IPCF, Intense Laser Irradiat Lab, I-56124 Pisa, Italy;
Ist Nazl Fis Nucl, Sez Pisa, Pisa, Italy;
Univ Pisa, Dipartimento Fis, Pisa, Italy