Ray-tracing simulations of a bent crystal X-ray optics for imaging using laser-plasma X-ray sources

Labate L., Galimberti M., Giulietti A., Giulietti D., Gizzi LA., Köster P., Laville S., Tomassini P.

Type
Papers in JCR/ISI journals
Year
2004
Journal
LASER AND PARTICLE BEAMS
Volume
22 (3)
Pages
253–259
Impact factor
0.575
Citations
12 (Web of Science, updated on 22 November 2026)
Keywords
plasma X-ray emission; ray-tracing simulations; X-ray imaging; X-ray optics;

INO authors

Giulietti Antonio, Giulietti Danilo, Gizzi Leonida Antonio, Labate Luca, Tomassini Paolo

Abstract

Ray-tracing simulations of an optical X-ray system based on a spherically bent crystal operating in Bragg configuration for monochromatic projection imaging of thin samples are presented, obtained using a code developed for that purpose. The code is particularly suited for characterizing experimental arrangements routinely used with laser-produced plasma X-ray sources. In particular, the spatial resolution of the imaging system was investigated and a careful study of the complex pattern of the X-ray backlighting beam was performed.

Affiliations

CNR - IPCF, Intense Laser Irradiat Lab, I-56100 Pisa, Italy;
Univ Bologna, Dipartimento Fis, Bologna, Italy;
Univ Pisa, Dipartimento Fis, Pisa, Italy;
INFM, Sez Pisa, Pisa, Italy