Resistively heated, high pressure, membrane and screw driven diamond anvil cell

Santoro M., Hajeb A., Gorelli F.

Type
Papers in JCR/ISI journals
Year
2020
Journal
HIGH PRESSURE RESEARCH
Volume
40 (3)
Pages
379–391
Impact factor
1.431
Citations
11 (Web of Science, updated on 22 November 2026)
Keywords
Diamond anvil cell; resistive heating; vacuum chamber; optical spectroscopy and X-ray diffraction

INO authors

Gorelli Federico Aiace, Santoro Mario

Abstract

High temperature is of paramount importance in high pressure science. One of the leading tools in this respect is the resistively heated diamond anvil cell (DAC), where the heat is provided by small heaters, positioned close to the diamond/gasket/sample region (internally heated DAC, IHDAC) or by wrapping the DAC body into bigger heaters (externally heated DAC, EHDAC). Although IHDACs can reach sample temperatures higher than 1000 K, they are difficult to handle and the heater/diamond/gasket/sample region may be affected by strong thermal gradients potentially hindering accurate temperature measurements. Here we present a novel EHDAC, which overcomes these issues by uniquely joining: (i) high mechanical precision for multi-Mbar, (ii) high temperature alloys for operating to 1000 K, (iii) membrane or screw driven, easily switchable between each other, (iv) operation into a vacuum chamber, (v) uniform temperature, (vi) facile handling, and (vii) possibility to add internal heaters for achieving even higher temperatures.

Affiliations

Ist Nazl Ottica, CNR INO, Via N Carrara 1, I-50019 Sesto Fiorentino, Italy; European Lab Non Linear Spect, LENS, Via N Carrara 1, I-50019 Sesto Fiorentino, Italy.

More information

This work was supported by Ministry of Education, Universities and Research, MIUR [grant number 2015HK93L7]. This work is also part of the SCENT project, which has received funding from the ATTRACT project funded by the European Union [grant number 777222].