A review of the current analytical approaches for evaluating, compensating and exploiting self-absorption in Laser Induced Breakdown Spectroscopy

Rezaei F., Cristoforetti G., Tognoni E., Legnaioli S., Palleschi V., Safi A.

Type
Papers in JCR/ISI journals
Year
2020
Journal
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
Volume
169
Pages
105878-1–105878-25
Impact factor
3.752
Citations
119 (Web of Science, updated on 22 November 2026)
Keywords
LIBS, Self-absorption, Spectroscopy, Quantitative analysis

INO authors

Cristoforetti Gabriele, Tognoni Elisabetta

Abstract

In this paper a review of different approaches for handling the self-absorption effect in Laser-Induced Breakdown Spectroscopy (LIBS) applications is presented. From an analytical point of view, self-absorption is a major issue affecting the accuracy of LIBS quantitative analysis. Self-absorption impacts emission line profiles and intensities, resulting in non-linear calibration curves and influencing the estimation of the sample composition. The main purpose of this work is presenting a summary of the experimental parameters giving raise to selfabsorption and a thorough study of the different strategies for its estimation, compensation and exploitation in order to improve the analytical accuracy of the LIBS technique.

Affiliations

Department of Physics, K. N. Toosi University of Technology, 15875-4416 Tehran, Iran; CNR-INO, Research Area of National Research Council, Via G. Moruzzi 1, Pisa, Italy; CNR-ICCOM, Research Area of National Research Council, Via G. Moruzzi, 1, Pisa, Italy; Laser and Plasma Research Institute, Shahid Beheshti University, G. C., Evin, Tehran, Iran