Flight qualification of a diode laser for path difference determination of a high-resolution Fourier transform spectrometer

Bianchini G., Lanfranchi M., Cortesi U.

Type
Papers in JCR/ISI journals
Year
2000
Journal
APPLIED OPTICS
Volume
39 (6)
Pages
962–965
Impact factor
1.359
Citations
12 (Web of Science, updated on 22 November 2026)
Keywords
Radiance; Spectroscopy; Spectral resolution

INO authors

Bianchini Giovanni

Abstract

Diode lasers offer a lightweight, rugged, and economic alternative to other types of laser source, but they often do not provide long-term stability and spectral purity of emission. We have developed a stabilized, near-infrared diode laser source that is capable of as much as 20 mW of single-mode output power for the effective replacement of a He-Ne laser in the path difference measuring system of a high-resolution (1.25-m maximum path difference) Fourier transform spectrometer. Laser characterization has been performed both in laboratory measurements and in flight tests, resulting in a relative frequency stability of better than 10-6 on a 6-h flight.

Affiliations

Istituto di Ricerca sulle Onde Elettromagnetiche, Consiglio Nazionale delle Ricerche, Via Panciatichi 64, Firenze, 50127, Italy