New approach to noise factor measurement of imaging devices

Anno: 1994

Autori: Ciamberlini C., Longobardi G., Ramazza P.L., Residori S.

Affiliazione autori: Istituto Nazionale di Ottica, Largo E. Fermi 6, 50125 Firenze, Italy

Abstract: A new method for noise factor determination in image intensifier tubes is described. The measurements made in second- and third-generation image intensifiers are in accordance with theory and experiments reported in the literature. This method allows one to carry out the measurement on the overall screen instead of the central area only.

Giornale/Rivista: OPTICAL ENGINEERING

Volume: 33(3)      Da Pagina: 845  A: 849

Parole chiavi: NOISE FACTOR; IMAGE INTENSIFIER TUBE; MICROCHANNEL PLATE; PHOSPHOR SCREEN;
DOI: 10.1117/12.157680

Citazioni: 4
dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2024-05-12
Riferimenti tratti da Isi Web of Knowledge: (solo abbonati)
Link per visualizzare la scheda su IsiWeb: Clicca qui
Link per visualizzare la citazioni su IsiWeb: Clicca qui