Design and characterisation of black-body sources for infrared wide-band Fourier transform spectroscopy

Anno: 2008

Autori: Palchetti L., Bianchini G., Castagnoli F.

Affiliazione autori: Istituto di Fisica Applicata Nello Carrara, IFAC-CNR, Via Madonna del Piano, 10, Sesto Fiorentino, 50019 Firenze, Italy

Abstract: The design and characterisation of small and cost-effective reference black-body sources for radiometric calibration of Fourier transform spectrometers in the mid and far infrared is presented. The paper describes the optimisation of the absorbing black-coating material and the study of two types of cavity geometry, one designed for collimated beams and one for focused beams. These two reference sources have been developed for the Radiation Explorer in the Far InfraRed spectrometer covering the 100-1400 cm-1 spectral range. They are currently used on-board two prototypes for ground and stratospheric balloon based atmospheric emission measurements. An emissivity of better than 0.9999 is obtained for both type of cavity geometry giving a calibration error, depending only on the temperature gradient inside the blackbody, of less than 0.5 K.

Giornale/Rivista: INFRARED PHYSICS & TECHNOLOGY

Volume: 51 (3)      Da Pagina: 207  A: 215

Parole chiavi: Black coatings; Cost effectiveness; Fourier transform infrared spectroscopy; Product design; Radiometry, Black coating materials; Blackbody radiation, Radiation
DOI: 10.1016/j.infrared.2007.06.001

Citazioni: 11
dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2024-04-21
Riferimenti tratti da Isi Web of Knowledge: (solo abbonati)
Link per visualizzare la scheda su IsiWeb: Clicca qui