Characterization and calibration of a variable-angle absolute reflectometer
Castellini C., Emiliani G., Masetti E., Poggi P., Polato P.P.
Autori INO
Castellini Carlo, Poggi Pasquale
Abstract
The instrument described is a newly designed reflectometer. Measurements are made with a polarized light beam as a function of wavelengt hand angle of incidence. Calibration tests have outlined the possibility of using the instrument for reflectance measurements on mirrors with good reproducibility (0.2%) and accuracy (better than 1%). This has been obtained by the alignment system of the sample and the absolute method used for the determination of its reflectance .
Affiliazioni
Comitato Nazionale per la Ricerca e lo Sviluppo dell'Energia Nucleare e delle Energie Alternative, S.P. Anguillarese Km. 1300-Casaccia-Roma, Italy; Istituto Nazionale di Ottica, Largo E. Fermi 6, 50125 Arcetri-Firenze, Italy; Stazione sperimentale del Vetro, 10 Via Briati, Murano-Venezia, Italy